Validation and
21 CFR Part 11 Compliance Certificate (ISO 13485 Software)
IMSXpress Document Control and QMS Management Software
|
Please download the current Validation and CFR Title 21 Part 11 Compliance Certificate form
this link:
|
PD-P11: 21 CFR Part 11 Compliance Traceability Matrix
|
Part 11 Requirement
|
IMSXpress Feature
|
11.10 Controls for closed systems
|
11.10 (a) Validation of systems to ensure accuracy, reliability, consistent intended
performance, and the ability to discern invalid or altered records.
|
IMSXpress software is validated to:
- IEC 62304:2006 Medical device software -Software life
cycle processes
- IEC TR 80002 Medical device software
- Guidance on the application of ISO 14971 to medical device software
- FDA Guidance for Industry 21 CFR Part
11; Electronic Records; Electronic Signatures - Validation, August 2001
|
11.10 (b) The ability to generate accurate and complete copies of records in both
human readable and electronic form suitable for inspection, review, and copying
by the agency.
|
Every document, record and attachment can be printed out or saved into an electronic
file. Audit trail reports can also be printed out or saved.
|
11.10 (c) Protection of records to enable their accurate and ready retrieval throughout
the records retention period.
|
Records are saved to a database on a server. Customer is responsible for insuring
that the database is regularly backed up; and for controlling access to the server.
All records are saved with a checksum (hash sum) to ensure integrity of records
and to protect records against tampering. Every time a record is opened or shown,
the checksum is recalculated and compared to the original checksum. If the checksums
don’t mach, a message is displayed.
|
11.10 (d) Limiting system access to authorized individuals.
|
Access to the IMSXpress system is controlled by log in user name and password. Following
password security policies can be set:
- Password length and complexity requirements
- Password expiration period (password aging)
- Preventing reuse of passwords
- Require changing password on first log-in
- Inactivity logout period
|
11.10 (e) Use of secure, computer-generated, time-stamped audit trails to independently
record the date and time of operator entries and actions that create, modify, or
delete electronic records. Record changes shall not obscure previously recorded
information. Such audit trail documentation shall be retained for a period at least
as long as that required for the subject electronic records and shall be available
for agency review and copying.
|
In IMSXpress records are never overwritten or deleted. A modified record is saved
as a new version of the record without overwriting the original record. Deleted
records are flagged ‘Deleted’ but are not actually deleted from the database. All
versions of a record and ‘Deleted’ records are identified with a time stamp and
an identification of the logged in user who created, modified or deleted the record.
The time stamp is the server time in a universal time format.
All record versions and ‘Deleted’ records can be viewed and printed out in the Audit
Trail module. Audit trail records are protected by checksums and cannot be altered
or deleted.
|
11.10 (f) Use of operational system checks to enforce permitted sequencing of steps
and events, as appropriate.
|
IMSXpress has built in operational system checks to enforce workflow steps that
preserve the integrity of electronically signed records. For example, all electronically
signed documents and records are sealed and cannot be modified after the signature
is executed. Another example is automatic, sequential numbering of records.
|
11.10 (g) Use of authority checks to ensure that only authorized individuals can
use the system, electronically sign a record, access the operation or computer system
input or output device, alter a record, or perform the operation at hand.
|
IMSXpress has the following security levels:
- System Administrator
- Module Administrator
- Authorization (use of electronic signature)
- Delete Record/Document
- Alter Record/Document
- Read Only
|
11.10 (h) Use of device (e.g., terminal) checks to determine, as appropriate, the
validity of the source of data input or operational instruction.
|
IMSXpress can only receive input from the keyboard and mouse device. The use of
keyboard and mouse is implicitly tested throughout the validation of the software.
IMSXpress has built in check routines and validation rules to ensure that data entered
into data fields is of correct type, length and/or range.
|
11.10 (i) Determination that persons who develop, maintain, or use electronic record/electronic
signature systems have the education, training, and experience to perform their
assigned tasks.
|
The use of electronic signatures is explained in the IMSXpress Tutorial. The tutorial
is integrated into the software and is immediately available to all users.
|
11.10 (j) The establishment of, and adherence to, written policies that hold individuals
accountable and responsible for actions initiated under their electronic signatures,
in order to deter record and signature falsification.
|
Customer is responsible for establishing and enforcing pertinent procedures.
|
11.10 (k)(1) Adequate controls over the distribution of, access to, and use of documentation
for system operation and maintenance.
|
Documentation for operation and maintenance IMSXpress software is integrated into
the software itself, and is thus immediately available to all users.
|
11.10 (k)(2) Revision and change control procedures to maintain an audit trail that
documents time-sequenced development and modification of systems documentation.
|
System documentation is integrated into the software itself and is developed, validated
and released as integral part of each software version.
|
11.30 Controls for open systems
|
Not applicable
|
11.50 Signature manifestations
|
11.50 (a) Signed electronic records shall contain information associated with the
signing that clearly indicates all of the following:
11.50 (a)(1) The printed name of the signer;
11.50 (a)(2) The date and time when the signature was executed; and
11.50 (a)(3) The meaning (such as review, approval, responsibility, or authorship)
associated with the signature.
|
Signed electronic records identify the signer name and position, the (server) date
and time when the signature was executed, and the meaning associated with the signature.
|
11.50 (b) The items identified in paragraphs (a)(1), (a)(2), and (a)(3) of this
section shall be subject to the same controls as for electronic records and shall
be included as part of any human readable form of the electronic record (such as
electronic display or printout).
|
The information required in 11.50(a) is controlled and protected in the same way
as the record itself (protected with a checksum, never overwritten and never deleted).
All this information is clearly displayed (printed) on the record itself.
|
11.70 Signature/record linking
|
11.70 Electronic signatures and handwritten signatures executed to electronic records
shall be linked to their respective electronic records to ensure that the signatures
cannot be excised, copied, or otherwise transferred to falsify an electronic record
by ordinary means.
|
The electronic signature information is embedded into the signed record itself (as
additional columns) and is protected by a common checksum. Any attempt to excise,
copy, or otherwise transfer the signature will invalidate the record checksum (hash
sum).
|
11.100 General requirements
|
11.100 (a) Each electronic signature shall be unique to one individual and shall
not be reused by, or reassigned to, anyone else.
|
Electronic signature consists of a signer code and password that the signer sets
up personally. The signer code and password are only known to the signer, and cannot
be reassigned or reused.
|
11.100 (b) Before an organization establishes, assigns, certifies, or otherwise
sanctions an individual’s electronic signature, or any element of such electronic
signature, the organization shall verify the identity of the individual.
|
Customer is responsible for compliance with this requirement.
|
11.100 (c) Persons using electronic signatures shall, prior to or at the time of
such use, certify to the agency that the electronic signatures in their system,
used on or after August 20, 1997, are intended to be the legally binding equivalent
of traditional handwritten signatures
|
Customer is responsible for compliance with this requirement.
|
11.200 Electronic signature components and controls
|
11.200 (a)(1) Employ at least two distinct identification components such as an
identification code and password.
(i) When an individual executes a series of signings during a single, continuous
period of controlled system access, the first signing shall be executed using all
electronic signature components; subsequent signings shall be executed using at
least one electronic signature component that is only executable by, and designed
to be used only by, the individual.
(ii) When an individual executes one or more signings not performed during a single,
continuous period of controlled system access, each signing shall be executed using
all of the electronic signature components.
|
IMSXpress employs two distinct identification components in the electronic signature:
signer code and password.
Every individual signing requires the use of both components.
|
11.200 (a)(2) Be used only by their genuine owners.
|
Signing code and password are personally set by the signature owner and are known
only to that individual.
|
11.200 (a)(3) Be administered and executed to ensure that attempted use of an individual's
electronic signature by anyone other than its genuine owner requires collaboration
of two or more individuals.
|
Signer code and password cannot be set by the System Administrator. To set a signer
code and password the user must be logged in to his or her IMSXpress account and
must know his or her log in name and password.
|
11.200 (b) Electronic signatures based upon biometrics shall be designed to ensure
that they cannot be used by anyone other than their genuine owners.
|
Not applicable
|
11.300 Controls for identification codes/passwords
|
11.300 (a) Maintaining the uniqueness of each combined identification code and password,
such that no two individuals have the same combination of identification code and
password.
|
Before accepting a new combination of signer code and password the system checks
whether the combination is unique.
|
11.300 (b) Ensuring that identification code and password issuances are periodically
checked, recalled, or revised (e.g., to cover such events as password aging).
|
Following signer password security policies can be set:
- Password length and complexity requirements
- Password expiration period (password aging)
- Preventing reuse of passwords
|
11.300 (c) Following loss management procedures to electronically deauthorize lost,
stolen, missing, or otherwise potentially compromised tokens, cards, and other devices
that bear or generate identification code or password information, and to issue
temporary or permanent replacements using suitable, rigorous controls.
|
System Administrators can disable electronic signing privilege for any user and
can reset the signer code and password. After the code and password are reset, the
user must set up new signing credentials. Users can also change their signing credentials
at any time.
|
11.300 (d) Use of transaction safeguards to prevent unauthorized use of passwords
and/or identification codes, and to detect and report in an immediate and urgent
manner any attempts at their unauthorized use to the system security unit, and,
as appropriate, to organizational management.
|
To use her electronic signature the user must be also logged in to her account.
Every electronic signing is traceable to:
- Windows/server user account
- IMSXpress user account
- Computer on which the signature was executed
- Server time stamp
|
11.300 (e) Initial and periodic testing of devices, such as tokens or cards, that
bear or generate identification code or password information to ensure that they
function properly and have not been altered in an unauthorized manner.
|
All software elements related to set up, changes, reset, etc. of electronic signature
signer code and password are formally validated for each software version release.
|